Test-retest repeatability and software reproducibility of myocardial flow measurements using rest/adenosine stress Rubidium-82 PET/CT with and without motion correction in healthy young volunteers

Research output: Contribution to journalJournal articleResearchpeer-review

Background
Myocardial flow reserve (MFR) assessment with cardiac positron emission computed tomography (PET/CT) is well established, and quantification relies on commercial software packages. However, for reliable use, repeatability and reproducibility are important. The aim of this study was therefore to investigate and compare between scans and software packages the repeatability and reproducibility of 82Rb-PET/CT estimated MFR.

Methods and Results
Forty healthy volunteers completed two 82Rb-PET/CT rest and adenosine stress scans. syngo.MBF (Siemens), quantitative-gated SPECT (QGS) (Cedars-Sinai), and Corridor4DM (4DM) were used for analyses. Motion correction was available for 4DM. Fifty percent were men and age was 24 ± 4 years (mean ± SD). Repeatability of MFR varied between scans. syngo.MBF: mean difference (95% CI) 0.26 (− 0.03 to 0.54), P = 0.07, limits of agreement (LoA): − 1.43 to 1.95; QGS: 0.19 (− 0.08 to 0.46), P = 0.15, LoA: − 1.38 to 1.76; 4DM: 0.08 (− 0.17 to 0.34), P = 0.50, LoA: − 1.37 to 1.53; and 4DM motion corrected: 0.17 (− 0.17 to 0.51), P = 0.32, LoA: − 1.89 to 2.22. MFR was higher using 4DM +/− motion correction compared with syngo.MBF and QGS (all P < 0.0001). Concordance between syngo.MBF and QGS was high (P = 0.83).

Conclusions
Reproducibility of MFR varied for the different software. The highest concordance between MFRs was found between syngo.MBF and QGS.
Original languageEnglish
JournalJournal of Nuclear Cardiology
Volume28
Pages (from-to)2860–2871
Number of pages12
ISSN1071-3581
DOIs
Publication statusPublished - 2021

    Research areas

  • Myocardial perfusion, Myocardial blood flow, Myocardial flow reserve, Positron emission tomography, Repeatability, Reproducibility